The following four tests can be used to assist in troubleshooting load cells. A precision power supply and multimeter are necessary.
Test 1: Mechanical Inspection
Each load cell should be inspected for physical damage. A thorough inspection of the entire length of any cables should be done. Nicked or abraded cables can short out a load cell. Inspect for distortion or cracks on all metal surfaces. Flexure surfaces, if any, should be parallel to each other and be perpendicular to both end surfaces. If the load cell has been physically deformed (bent, stretched, or compressed from original configuration) it is non-repairable. Consult Measurement Specialists for load cell application information.
Test 2: Zero Balance (electrical output with no load)
Changes in the zero balance are usually caused by residual stress in the sensing area. Residual stresses result from overloading the cell, and /or from millions of cycles.
With a millivoltmeter, measure the load cell's output under a 'no load' condition, (no weight on the load cell). The output of a trimmed cell should typically be within +/- .1% of rated output. Always consult published specifications to obtain the correct values.
EXAMPLE: If the typical load cell output =2 mV/V and the typical power supply output =10V, then the full scale output would be =20 mV. Typical zero balance =+/-.1% of rated output. The zero reading should then be within +/- .02 mV.
If the output is greater than the zero balance tolerance, the cell is damaged but may still be useable. When the zero reading exceeds the zero balance tolerance but is less than 50% of the full scale output, the load cell may be electronically 're-zeroed' if the electronic readout with the cell allows it; however, there is no guarantee that the load cell will remain linear. If a load cell has more than 50% full scale zero balance shift, it may require 're-gauging'.
Test 3: Bridge Resistance
Out of tolerance bridge resistance values are usually caused by the failure of a compensating element. Another cause is a broken or burned bridge wire. These failures commonly result from electrical transients or lightening strikes.
TO TEST FOR BRIDGE RESISTANCE: Measure the resistance across each pair of input and output leads. The input and output resistance is typically 350 +/- 3.5 ohms, however, there are many variations, depending on each load cell's published specifications. Consult the published specifications of the subject load cell for proper values and tolerances. If these readings are 'out of spec', the load cell requires repair.
Test 4: Resistance to Ground
Current flow from the load cell's electronics to the load cell body is usually caused by water contamination within the load cell or its cable.
TO TEST FOR LEAKAGE: Connect all of the input, output, sense and ground leads together and measure the resistance between the load cell body and the leads with a megohmmeter. Make sure good contact is made with the metal load cell sensor. The reading should be at least 5000 Megohms. If the load cell fails this test, retest without the ground wire. If the load cell passes this test, a problem may exist with the load cell cable. If the test fails the load cell requires repair.
- Courtesy of Measurement Specialists
Test 1: Mechanical Inspection
Each load cell should be inspected for physical damage. A thorough inspection of the entire length of any cables should be done. Nicked or abraded cables can short out a load cell. Inspect for distortion or cracks on all metal surfaces. Flexure surfaces, if any, should be parallel to each other and be perpendicular to both end surfaces. If the load cell has been physically deformed (bent, stretched, or compressed from original configuration) it is non-repairable. Consult Measurement Specialists for load cell application information.
Test 2: Zero Balance (electrical output with no load)
Changes in the zero balance are usually caused by residual stress in the sensing area. Residual stresses result from overloading the cell, and /or from millions of cycles.
With a millivoltmeter, measure the load cell's output under a 'no load' condition, (no weight on the load cell). The output of a trimmed cell should typically be within +/- .1% of rated output. Always consult published specifications to obtain the correct values.
EXAMPLE: If the typical load cell output =2 mV/V and the typical power supply output =10V, then the full scale output would be =20 mV. Typical zero balance =+/-.1% of rated output. The zero reading should then be within +/- .02 mV.
If the output is greater than the zero balance tolerance, the cell is damaged but may still be useable. When the zero reading exceeds the zero balance tolerance but is less than 50% of the full scale output, the load cell may be electronically 're-zeroed' if the electronic readout with the cell allows it; however, there is no guarantee that the load cell will remain linear. If a load cell has more than 50% full scale zero balance shift, it may require 're-gauging'.
Test 3: Bridge Resistance
Out of tolerance bridge resistance values are usually caused by the failure of a compensating element. Another cause is a broken or burned bridge wire. These failures commonly result from electrical transients or lightening strikes.
TO TEST FOR BRIDGE RESISTANCE: Measure the resistance across each pair of input and output leads. The input and output resistance is typically 350 +/- 3.5 ohms, however, there are many variations, depending on each load cell's published specifications. Consult the published specifications of the subject load cell for proper values and tolerances. If these readings are 'out of spec', the load cell requires repair.
Test 4: Resistance to Ground
Current flow from the load cell's electronics to the load cell body is usually caused by water contamination within the load cell or its cable.
TO TEST FOR LEAKAGE: Connect all of the input, output, sense and ground leads together and measure the resistance between the load cell body and the leads with a megohmmeter. Make sure good contact is made with the metal load cell sensor. The reading should be at least 5000 Megohms. If the load cell fails this test, retest without the ground wire. If the load cell passes this test, a problem may exist with the load cell cable. If the test fails the load cell requires repair.
- Courtesy of Measurement Specialists
+ comments + 4 comments
Ultra Precision strain gauge based ‘S’ beam Load Cell is versatile, compact and Rugged. It is used for tension measurement, weighing, industrial automation, and research application. It is universal Load Cell Suitable for both tension and compression modes.
load cells
very nice blogs your site is good.have shared useful information ......Thankq for load cell amplifier
load cells
Product B6G Load Cell Product loadcell
Stainless steel IP66 single point loadcell.
Suitable for single or double load cells structured platform scales.
Recommended platform size:600x600 mm
These are such useful ideas for load cells! I'm thinking of getting a stainless steel load cell. What can you tell me about a 1000 kg capacity?
Celine | http://www.nuweigh.com.au/load_cells/